PGU-OEM H2

  • 2 Channels: 1: 100 mA, 2: 2000 mA
  • specifically designed for hydrogen experiments with its own measurement method
  • Interface and Software EcmWin included

PGU-OEM H2 is a special version for permeation experiments in combination with our devanathancell. The device consits of two modules, one for the oxidation side and one for the charging side. Both modules are double galvanically isolated and can communicate with each other via EcmWin.
The idea of this design is the double galvanic isolation of the modules and the low noise setup by reducing cables and other parts which can catch disturbance from the environment. This is very important for hydrogen permeation measurement. Any leakage current between the devices on additional ways than via the electrochemical cell distorts the measurement result. By testing materials that are high resistant against hydrogen permeation for example, the expected permeation current is in a range of 5 to 10nA. Any leakage current could be higher if it is present.

Type

Potentiostat/Galvanostat

Electrode connections

CE, RE, WE, WE-SENSE

Polarization range

Module 1: ± 10V / ± 2000 mA

Module 2: ± 10V / ± 100 mA

Current range

Module 1: 10 ranges to 10 nA

Module 2: 10 ranges to 100 pA

Compliance voltage

± 13 V

Resolution

Module 1: 0.2 %

Module 2: 0.1 pA = 1 mV

Methods (via EcmWin)

OCP, Hold (Constant), Scan, CV, Pulse, Data and Script measurement, H2-Method: interaction of oxidation side with charging side

Interfaces

Ethernet/USB

Measurement data acquisition

24 Bit A/D converter

Scanner

26 Bit (step size 330 nV)

Impedance analyzer

(optional)

Frequency range 10 µHz-100 kHz

Scroll to Top